Minitest 650FN – The Robust coating thickness measurement gauge
Available on backorder
MiniTest 650 is a robust gauge to measure coating thickness on rough and abrasive surfaces. The video explains how to calibrate correctly in order to achieve highest precision measuring non-magnetic coatings on steel and insulating coatings on non-ferrous metals.
Price from: € 1,350.0
(Price just for reference only)
Available on backorder
MiniTest 650 – The Robust coating thickness measurement gauge. The MiniTest 650 has a fixed external probe for all non-magnetic coatings on steel and all insulating coatings on nonferrous metals. For measurement on rough steel surfaces, e.g. in shipyards, the F probe comes equipped with special wear protection to ensure long service.
MiniTest 650 FN – coating thickness measurement gauge
To measure non-magnetic coatings (varnishes, enamels etc.) on ferromagnetic substrates (e.g. steel used in construction and for auto body) and to measure insulating coatings (varnishes, enamels, anodic coatings etc.) on non-ferrous metals (aluminium, copper, brass etc.).
With statistics display (number of readings taken, mean values, standard deviation, maximum, minimum).
Measuring range 0 to 2 mm.
Interchangeable measuring units µm or mils.
Enables 1-point and 2-point calibrations.
Non-destructive coating thickness measurement:
- Non-magnetic coatings like varnishes, enamels, chrome and zinc on ferromagnetic base materials (e.g. steel)
- Insulating coatings like varnishes and anodising aluminium on non-ferrous metals (e.g. aluminium, copper, die-cast zinc)
Measuring principles
- Magnetic induction (F-type sensors) according to DIN EN ESO 2808, method 7C
- Eddy current (N-type sensors) according to DIN EN ESO 2808, method 7D
Features
- Coating thickness measurement of up to 2 and 3 mm
- Universal probe offering automatic substrate identification for quick measurement on steel and non-ferrous metals
- Wear-resistant probe tips with practically unlimited durability
- USB-interface to transfer data to a PC
- Statistical evaluation of the readings (measurement data)
Supply schedule
- Gauge MiniTest 650 with fixed probe
- 3 batteries AAA / LR03 / Micro
- Control plate(s), calibration standards in folding case
- Operating manual on a CD
- Quick reference guide
- Factory certificate
- Soft case
Minitest 650FN – Technical details
- Measuring range: FN (dual sensor): 0 µm – 2000 µm / 80 mils
- Measuring uncertainty: ± (2 % of reading + 2 μm / 0.08 mils)
- Minimum curvature radius: 5 mm / 0.2“ convex. 25 mm /1“ concave
- Minimum measuring area: Ø 20 mm
- Minimum base thickness: Sensor F: 0.5 mm. Sensor N: 50 µm / 2 mils
- Display: 4-digit screen data (11 mm / 0.44”)
- Measuring units: μm – mils, user selectable
- Calibration: standard, 1-point and 2-point calibration
- Statistics: calculated from maximum 9.999 readings – mean value – standard deviation – number of readings – minimum – maximum
- Interface: USB
- Power supply: 3 Micro AAA batteries (for more than 10,000 readings)
- Dimensions: gauge: 70 mm x 122 mm x 32 mm. sensor: Ø 15 mm x 62 mm
- Weight: 225g
- Ambient temperature: gauge: 0 °C – 50 °C. sensor: -10 °C – 70 °C
Maker | |
---|---|
Origin | Made in Germany |